Updated: Jun-23-2020

General Information

Type of service provider
Calibration laboratory
Physical postal address

P.O Box 89465-80100
Mombasa
Kenya

Phone number
0732 424 320
E-mail
msapetrogas [at] gmail.com
info [at] mombasapetrogas.co.ke
Has the laboratory been accredited in any scopes?
No
Does the laboratory have other sites?
No

Personnel

Chief Executive Officer (CEO) / Head of Organization

CEO Name
Austin Wise Katuu
CEO Phone
CEO E-mail
austin.katuu [at] mombasapetrogas.co.ke

Total laboratory personnel

Total laboratory personnel
9
Male laboratory personnel
8
Female laboratory personnel
1

Technical personnel

Technical personnel total
7
Male technical personnel
7

Technical Information

Fields of Metrology

Flow
Pressure
Volume and density
Scope of calibration services offered
Calibration object
Horizontal above ground tanks

Underground tanks
Quantity/ Property
Volume
Procedure reference
Tank calibration procedure
MPSL/P/102
Range/ Limit
0.5-50,000L
CMC
50,000 +/- 30L
Accreditation status
No
Calibration object
Positive displacement flow meters
Micro motion meters
Accuload meters
Quantity/ Property
Flow
Procedure reference
Meter calibration
Procedure
MPSL/P/101
Range/ Limit
15-2000L/Min
CMC
Q min = 15 +/- 0.03 L/min

Qr = 150 +/- 0.4 L/min

Qmax= 2000+/- 4L/min
Accreditation status
No
Calibration object
Industrial pressure gauges

Bourdon tube pressure gauges
Quantity/ Property
Pressure
Procedure reference
Test gauge method

MPSL/P/103

Range/ Limit
0.1-600bar

1-10kg/cm2
CMC
0.1-600bar +/- 0.5%

1-10kg/cm2 +/- 0.8%
Accreditation status
No
Uncertainty of measurements reports
Yes
Statement identifying how the measurements are metrologically traceable
Yes

How does the laboratory ensure international traceability of its measurements standards?

Through calibration certificates issues by an accredited national laboratory
Yes
Through calibration certificates issues by an accredited overseas laboratory
No
Does the lab participate in any inter- laboratory comparison?
Yes
Please provide details about the scheme
The scheme is organised by an NMI (KEBS) in our case they perform the initial and final measurements of an item. The other measurements are taken by the other participating labs including ours.
Does the laboratory sub-contract calibration work to other laboratories?
No

Quality Control / Quality Management

Does the lab operate an internal quality control (QC) system to ensure quality of measurement?
Yes
Does the lab operate a management system (or Quality Management System) for assuring the quality of calibrations?
Yes
Is the system documented in a Quality Manual?
Yes